| Model Number |
40000-07T |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported two sec 20dp, texium & hanger v/nv had component separation issues.The following information was provided by the initial reporter: "i have 2 more texium secondary sets that came apart like the others.".
|
| |
|
Additional Manufacturer Narrative
|
|
H.6.Investigation: the customer reported the product came apart and returned two used samples of material #40000-07t.The samples were clearly separated below the drip chamber, and the complaint is verified.The root cause for the separation is insufficient solvent applied during assembly.This issue has been ongoing and manufacturing has made corrective actions to address the issue.These include a project to update the solvent applicator optimal screw depth specification and preparation, as well as balancing the manufacturing cell operations.A device history record review could not be performed on model 40000-07t because a valid lot number was not provided by the customer.H3 other text : see h.10.
|
| |
|
Event or Problem Description
|
|
It was reported two sec 20dp, texium & hanger v/nv had component separation issues.The following information was provided by the initial reporter: "i have 2 more texium secondary sets that came apart like the others.".
|
| |
|