| Model Number |
71654502 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/07/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an internal fixation a std drill guide broke while inserting nail.Surgery was finished with a smith and nephew backup device, with 5 minute of surgical delay.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).The device, used in treatment, was returned for evaluation.A visual inspection of the standard drill guide confirms the threads are damaged.Also there are multiple scratches, burrs and gouges in the metal.There are no confirmed broken pieces on the drill guide.The device shows signs of significant wear and use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The device shows significant signs of wear/usage.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|