• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. STD DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. STD DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71654502
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that, during an internal fixation a std drill guide broke while inserting nail.Surgery was finished with a smith and nephew backup device, with 5 minute of surgical delay.Patient was not injured as consequence of this problem.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The device, used in treatment, was returned for evaluation.A visual inspection of the standard drill guide confirms the threads are damaged.Also there are multiple scratches, burrs and gouges in the metal.There are no confirmed broken pieces on the drill guide.The device shows signs of significant wear and use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The device shows significant signs of wear/usage.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STD DRILL GUIDE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12542698
Report Number1020279-2021-07258
Device Sequence Number1508008
Product Code HSB
UDI-Device Identifier03596010585837
UDI-Public03596010585837
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Model Number71654502
Device Catalogue Number71654502
Device Lot Number10FSP0015
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2021
Initial Date Received by Manufacturer 09/07/2021
Supplement Date Received by Manufacturer10/21/2021
Initial Report FDA Received Date09/28/2021
Supplement Report FDA Received Date10/22/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-