| Model Number |
66800472 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the foot pedal cord of a versajet ii multifunction footswitch is broken at the strain relief of the pedal end, it needs to be repaired or exchange the foot pedal.As this happened in a non-therapeutic environment, there was not patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the device, intended for use in treatment, was returned for evaluation.The visual inspection of the returned device noted bare wiring on the cord, and the functional evaluation revealed that the footswitch would not activate the console, establishing a relationship between the device and the reported event.The root cause was identified as wear and tear.The device should never be used when, bare wires are exposed.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.A complaint history review found further instances of the reported events.This investigation is now complete with no further action deemed necessary.Smith + nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|