SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 3-4 9MM RT; PROST, KNEE, PATFEMTIB, SEMI-CNSTRND, CEMENTED, POLY/MET/POLY
|
Back to Search Results |
|
| Model Number |
74035231 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during total knee arthroplasty, a journey ii bcs articular insert trial size 3-4 9 mm right broke in half outside the patient.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device, intended for use in treatment was returned for evaluation.A visual inspection of the returned device confirmed the stated failure mode.The journey ii bcs art isrt trl sz 3-4 9mm rt is cracked in half rendering the device inoperative.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|