| Model Number |
DWD183 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
|
| |
|
Event or Problem Description
|
|
The customer reported that the glenosphere would not attach onto baseplate.A different glenopshere was used.This was a different sized implant and the surgeon had to use a different humeral component to match.The same base plate was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction h6 method code.The reported event could not be confirmed since the device was not returned for evaluation and no other evidence was provided for evaluation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
The customer reported that the glenosphere would not attach onto baseplate.A different glenopshere was used.This was a different sized implant and the surgeon had to use a different humeral component to match.The same base plate was used.
|
| |
|