Model Number 71654502
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during an internal fixation surgery, a trigen standard drill guide broke, no pieces fell inside the patient.Surgery was resumed, without any delay, with a back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the standard drill guide confirms the threads are damaged.Also there are multiple scratches, burrs and gouges in the metal.There are no confirmed broken pieces on the drill guide.The device shows signs of significant wear and use.A review of complaint history did not reveal additional complaints for the listed batch for the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.The device shows significant signs of wear/usage.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STD DRILL GUIDE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12544863
MDR Text Key273767428
Report Number1020279-2021-07287
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010585837
UDI-Public03596010585837
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71654502
Device Catalogue Number71654502
Device Lot Number17KSP0002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2021
Initial Date Manufacturer Received 09/13/2021
Initial Date FDA Received09/28/2021
Supplement Dates Manufacturer Received10/25/2021
Supplement Dates FDA Received10/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-