Model Number M3536A
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2021
Event Type  malfunction  
Event Description
It was reported to philips that the device had bent pins.There was no patient involvement.
 
Manufacturer Narrative
It was reported to philips that the device had bent pins.The device was received by bench for repair on 12/07/2021.The noted failure was identified during inspection of the device by an authorized bench technician.As a result of the issue, it was determined that the battery pca would need to be replaced and defibrillator energy is low and therapy capacitor would need to be replaced.Defibrillator issue carried out on pr# (b)(4).Upon conclusion of the evaluation, it was determined that this was a malfunction of the battery pca.The battery pca was replaced to resolve the reported issue and the device was scheduled to be returned to the customer site.No further evaluation is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX -EMS DEFIBRILLATOR
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
lois price
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key12546880
MDR Text Key273781792
Report Number3030677-2021-14581
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838006652
UDI-Public00884838006652
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3536A
Device Catalogue NumberM3536A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-