| Model Number |
709031 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ref.Id: (b)(4).The investigation is still ongoing on this event.When the investigation is completed a follow-up report will be sent to the fda.
|
| |
|
Event or Problem Description
|
|
The customer reported that the current image is mixed with the previous image.
|
| |
|
Additional Manufacturer Narrative
|
|
Ref.Id:(b)(4).The combidiagnost system supports general radiography and fluoroscopy imaging.Digital images can be archived in external digital archives (pacs).The pacs system administrator is responsible for the proper import and display of such digital images from the connected acquisition systems.The issue was caused in the pacs so outside the philips acquisition system.No further information was provided as to the type of examination or how the images were mixed up.The hospital pacs system administrator solved the issue in the pacs.The event was originally reported to the authorities as not enough information was available to make a definite reportability decision.Since that time, additional information was received which revealed that the event does not meet the criteria for reporting.Correction: h6 results and conclusion code.
|
| |
|