| Model Number |
1012449-20 |
| Medical Device Problem Codes |
Material Separation (1562); Deformation Due to Compressive Stress (2889)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported kink on the shaft and strain relief were confirmed.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported complaints appear to be related to operational context.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported during preparation of the 3.0x20mm nc trek balloon dilatation catheter, the proximal shaft was kinked.Another unspecified balloon was used to successfully complete the procedure.There was no patient involvement and no clinically significant delay.No additional information was provided.Per device analysis the hypotube was separated 5mm proximal to the strain relief distal end.The account could not confirm when the separation occurred.
|
| |
|