Medtronic received a report that the one pipeline experienced resistance in the hub of the phenom 27 microcatheter, two pipelines deployed in the catheter hub, and a third failed to open distally and broke during retrieval.The patient was undergoing flow diversion treatment for an unruptured, saccular aneurysm located in the left ica.The max diameter was 2.9mm, and the neck diameter was 1.5mm.The patient's vessel tortuosity was normal.The landing zone was 3.5mm distal and 4.5mm proximal.The access vessel was the left ica, which was 3.5-4.5mm in diameter.It was reported that the first pipeline was inserted into the first phenom 27 catheter and deployed in the hub during insertion.The device was removed.A second pipeline was inserted into the catheter with extra precaution.This device had resistance in the catheter hub and also deployed in hub.The tip of the sheath had been seated securely in the hub.The second pipeline was removed along with phenom 27 catheter which was replaced with a new phenom 27.A third pipeline was inserted into the catheter and was able to be introduced into catheter.During deployment of this pipeline, the device would not open distally and had to be retrieved.The middle part of the pipeline had been positioned in a bend, and more than 50% had been deployed when it failed to open.The pipeline had been resheathed 2 or less times, and no further steps were taken to open the device.The third pipeline was removed back into introducer hub; however, unknown at the time, the 4.75x12 pipeline had broken at the proximal bumper during retrieval.A fourth pipeline was introduced into the catheter.As device was being advanced, fluoro showed the tip of the third pipeline still inside the distal end of the phenom 27.The fourth pipeline 4.25x12 was removed, and the phenom 27 was then removed with the third 4.75x12 pipeline still inside.A new phenom 27 was then placed back in position.A fifth pipeline (4.25x10) was then introduced into catheter.This device was able to be deployed safely and angiographically looked well apposed with good aneurysm stasis. the patient did not experience any injury or complications.The devices were prepared and flushed according to the instructions for use (ifu).Ancillary devices include a penumbra 8fr neuron max sheath, navien 058 guide catheter, and phenox portal 014 guidewire.Additional information received reported that there was no damage observed to the catheter or pushwire.
|
|
Product analysis as found condition: the pipeline flex braid was returned for analysis within a shipping box, within a tyvek bio-pouch, and within primary and secondary plastic bio-pouches.Visual inspection/damage location details: the pipeline flex braid was returned already detached from the pusher; therefore, the braid ends (proximal, distal) could not be identified.The pipeline flex braid was found damaged.Testing/analysis: none conclusion: based on the device analysis and reported information, the customer¿s ¿device opens prematurely¿ report could not be confirmed.Premature deployment within the catheter hub could be caused by use of an improper hubbing technique (under tightening of the rhv/sheath not firmly seated into hub).However, improper hubbing technique could not be ruled out as a potential root cause for the event.If information is provided in the future, a supplemental report will be issued.
|