The event involved a plum set where a hole was found on the tubing which resulted in a leak, 10 minutes after infusion of nicardipine, for high blood pressure treatment.There was no unprotected exposure to the drug leak.There was patient involvement but no patient harm.No additional information was provided.
|
Received an image from the customer showing the set, with fluid, inside of a bag.No damage or anomalies were evident from the image.Received one used, list #146870489 (primary plum set, clave secondary port, clave y-site, secure lock, 103 inch; lot #5192799) and one used needle (list #unknown, lot #unknown).The set was visually examined and a large cut was found in the tubing.The set was air pressure leak tested and the aforementioned cut was the only leak site.The cut was examined and the edges were clean.No packaging was returned for evaluation.The complaint can be confirmed, the probable cause is due to contact with a sharp instrument during use.The manufacturing process was reviewed and it was determined that this type of damage would not have occurred during manufacturing.There are no sharp objects or instruments on the manufacturing floor capable of the observed damage as a preventative measure.The lot history was reviewed and no nonconformities were identified that may have contributed to the reported complaint.
|