| Model Number |
1012278-08 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record and complaint history could not be conducted because the lot number was not provided.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during a right coronary artery (rca) intervention a non-abbott device fractured in the vessel and the catheter tip could not be snared.A trek 4.0x8mm balloon dilatation catheter was advanced, but had some difficulty passing the retained catheter tip in the rca.Once at the desired location, the balloon was inflated to 16 atmospheres and then to 14 atmospheres for 12 seconds in the mid rca at the site of the broken catheter tip before the balloon burst.The device was removed without issue.There was no reported adverse patient effect and no clinically significant delay in the procedure.Another balloon was used to continue the procedure.No additional information regarding this issue was provided.
|
| |
|