• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL AMPERE¿ RF ABLATION GENERATOR; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number H700488
Medical Device Problem Codes Device Emits Odor (1425); Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
During a re-do atrial fibrillation procedure, a procedure delay occurred.The ampere suddenly shutdown while tachycardia was induced after ablation a burning smell was noted from the ampere.The power cable was replaced, but the issue persisted.The procedure was delayed one hour until a replacement unit was able to be found and the procedure was completed with no adverse patient consequences.
 
Additional Manufacturer Narrative
One ampere¿ rf ablation generator was received into the lab for analysis.An inspection of all internal components was performed prior to testing due to the field reported issue.Ac power was then applied to returned device which failed to initialize, confirming the reported electrical issue.No discolored internal components were observed, and no odors were noted.Additional troubleshooting isolated the root cause to abnormal functionality of the power supply and rf/mn printed circuit boards.Based on the information provided to abbott and the investigation performed, the reported event was confirmed and the root cause of the field reported issue was isolated to abnormal functionality of the power supply and radio frequency/ mn boards.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPERE¿ RF ABLATION GENERATOR
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12553634
Report Number2184149-2021-00273
Device Sequence Number19156374
Product Code LPB
UDI-Device Identifier05415067004888
UDI-Public05415067004888
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110016
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH700488
Device Catalogue NumberH700488
Device Lot Number5509649
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/07/2021
Supplement Date Received by Manufacturer11/05/2021
Initial Report FDA Received Date09/30/2021
Supplement Report FDA Received Date11/08/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/23/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-