• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFCSIGMA POST FEMAUG TRL4MM S4; KNEE INSTRUMENT : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFCSIGMA POST FEMAUG TRL4MM S4; KNEE INSTRUMENT : FEMORAL TRIALS Back to Search Results
Model Number 96-1886
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Two damaged items at (b)(6) hospital: damaged posterior augment ref (b)(4) ¿ no longer sits in femoral trial securely.Worn femoral punch ref (b)(4) ¿ distal tip is mis-shapen and worn down and no longer fit for purpose.Identified by (b)(6).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device was received for examination.Therefore, the reported event could not be confirmed.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) was not possible, because the required lot code was not provided.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative:  added: b3, b5, h6 (medical device problem code).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that a small plug on augment did not sit securely into the femoral trial and seemed loose or more susceptible to falling off.The "v" of the extractor tip was worn down and disfigured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFCSIGMA POST FEMAUG TRL4MM S4
Common Device Name
KNEE INSTRUMENT : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12554267
Report Number1818910-2021-21549
Device Sequence Number4080340
Product Code HWT
UDI-Device Identifier10603295240396
UDI-Public10603295240396
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96-1886
Device Catalogue Number961886
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/23/2021
Supplement Date Received by Manufacturer09/30/2021
10/25/2021
Initial Report FDA Received Date09/30/2021
Supplement Report FDA Received Date10/05/2021
10/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-