• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC GUIDE BLOCK FOR TWO-COLUMN PLATE/NARROW 6H HD/LT; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC GUIDE BLOCK FOR TWO-COLUMN PLATE/NARROW 6H HD/LT; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number 03.111.501
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6), 2021, patient underwent a procedure for distal radius.The 2.4mm cortex screws in the va distal radius set don¿t fit through the distal drill guides and the screw instead locks the guide down to the bone.While examining the screw head it¿s visible to see a slight difference between ones that fit and the ones that don¿t.The surgeon tried to use distal drill guide blocks and the screw wouldn¿t fit through block which locks the plate to the radius.Procedure was completed successfully without any surgical delay and no patient consequences reported.Concomitant medical products: unk - plates: va-lcp distal radius (part# unknown; lot# unknown; quantity: 1).This report is for one (1) guide block for two-column plate/narrow 6h hd/lt.This is report 2 of 20 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE BLOCK FOR TWO-COLUMN PLATE/NARROW 6H HD/LT
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12555038
Report Number2939274-2021-05810
Device Sequence Number13671151
Product Code FZX
UDI-Device Identifier10886982072436
UDI-Public(01)10886982072436
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.111.501
Device Catalogue Number03.111.501
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/01/2021
Supplement Date Received by Manufacturer10/28/2021
Initial Report FDA Received Date09/30/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.; 2.4 CORTEX SCREW SLF-TPNG T8 SD REC 14.
-
-