It was reported that balloon rupture, stent deployment issue and shaft break occurred, and the patient experienced discomfort and hematoma.The 90% stenosed target lesion was located in the mildly tortuous and severely calcified left anterior descending artery.A 3.00 x 38mm synergy xd drug-eluting stent was deployed for treatment.However, during the first inflation at 6 atmospheres, the balloon ruptured leaving behind partially deployed stent.The physician needed to recross stent and crush the stent against the side wall with additional stent.A shaft break was also noted 30mm distal from the tip.Furthermore, the patient experienced discomfort and hematoma.The stent delivery system was removed and the procedure was completed with a 3.0x38 non-boston scientific stent.No further patient complications were reported and the patient was stable post procedure.
|