• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS ANGIOSCULPT OTW PTA; CATHETER, PTA, CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS ANGIOSCULPT OTW PTA; CATHETER, PTA, CUTTING/SCORING Back to Search Results
Model Number 2076-6040
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient's dob or age at time of event, gender, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data or medical history are unknown.This information was not available from the facility.Report source: foreign- (b)(6).The angiosculpt device has not been returned by the facility, thus no returned product investigation was performed.
 
Event or Problem Description
The angiosculpt device was used to treat a peripheral lesion.During the first inflation at 4-6 atm, the balloon ruptured.The procedure was completed with a new angiosculpt device.No patient injury reported.This product problem is being submitted conservatively because the balloon ruptured below rbp.
 
Additional Manufacturer Narrative
Block d9/g3/h3: the angiosculpt device was returned for evaluation.Block h3: visual inspection found an elevated scoring element.During functional testing using an indeflator filled with green color dye water, the device was pressurized and at less than 2 atm, a pinhole leak on the balloon was present.Block h6: the ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT OTW PTA
Common Device Name
CATHETER, PTA, CUTTING/SCORING
Manufacturer (Section D)
SPECTRANETICS
5055 brandin court
fremont CA 94538
MDR Report Key12555768
Report Number3005462046-2021-00056
Device Sequence Number17167513
Product Code PNO
UDI-Device Identifier00813132021399
UDI-Public00813132021399
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K100303
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/09/2023
Device Model Number2076-6040
Device Catalogue Number2076-6040
Device Lot NumberG20110019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/08/2021
Supplement Date Received by Manufacturer10/04/2021
Initial Report FDA Received Date09/30/2021
Supplement Report FDA Received Date10/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK MFG AND SIZE: GUIDE CATHETER; UNK MFG AND SIZE: GUIDE WIRE; UNK MFG AND SIZE: INTRODUCER SHEATH
-
-