| Model Number |
8400D |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One unit was returned for investigation.There were lcd lines missing, with a lens crack and co2 pump was shifted.All these signs pointed towards the device suffering an impact such as a drop.Root cause traced to user.No dhr review done.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken lcd and calibration issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Please disregard the initial report submitted with the mfr number: 3012307300-2021-09699.The report was submitted in error.
|
| |
|