Model Number 8400D
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Calibration Problem (2890)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One unit was returned for investigation.There were lcd lines missing, with a lens crack and co2 pump was shifted.All these signs pointed towards the device suffering an impact such as a drop.Root cause traced to user.No dhr review done.
 
Event or Problem Description
It was reported that there was a broken lcd and calibration issues.
 
Additional Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Please disregard the initial report submitted with the mfr number: 3012307300-2021-09699.The report was submitted in error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI
Common Device Name
CARBON DIOXIDE GAS ANALYZER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12555884
Report Number3012307300-2021-09699
Device Sequence Number17141622
Product Code CCK
UDI-Device Identifier10610586036736
UDI-Public10610586036736
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K991086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8400D
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2021
Supplement Date Received by Manufacturer05/02/2023
Initial Report FDA Received Date09/30/2021
Supplement Report FDA Received Date05/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-