| Model Number |
7015 |
| Medical Device Problem Codes |
Contamination (1120); Unsealed Device Packaging (1444)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device inner package was compromised.A 3.0mm x 15mm quantum maverick balloon catheter selected for use.However, after opening the outer package, it was noticed that the sticking place of the inner package was opened and exposed to air.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
|
| |
|
Event or Problem Description
|
|
It was reported that device contamination might occurred.The target lesion was located in the left anterior descending artery.A 3.0mm x 15mm quantum maverick balloon catheter was advanced for dillatation.However, the device inner package was opened and exposed to air.The procedure was completed with a different device.There were no patient complications reported and the patient status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device codes corrected.The returned product consisted of an nc emerge balloon catheter and its tyvek inner packaging.No outer box was returned.Photo media was also provided.The product pouch and tyvek seal were visually inspected.The bottom tyvek (white) seal was pulled and curled up on the back flap of the packaging far enough that the device was elementally exposed.However, all seals on the package were complete and sealed at one point without error.Inspection of the remainder of the device, apart from the observed damage to packaging, revealed no other damage or irregularities.Photo media was reviewed and found to match analysis of the device.
|
| |
|
Search Alerts/Recalls
|