• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; 2V-4 3-0 PDO 24 X 24

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; 2V-4 3-0 PDO 24 X 24 Back to Search Results
Model Number SXPD2B426
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem Post Operative Wound Infection (2446)
Event Date 08/18/2021
Event Type  Injury  
Manufacturer Narrative
A batch review of the finished good lot and components confirmed there were no quality issues noted throughout the incoming inspection, manufacturing, sterilization, in-process, or final inspection process.To date samples or photos have not been returned for review and/or analysis.Without receiving details of the procedure, placement of the device, culture results and patient health history, a definitive root cause for the reported event of infection cannot be confirmed with certainty.Adverse effects associated with the use of this device may include, wound dehiscence, failure to provide adequate wound support in closure of the site where expansion, stretching or distension occur, failure to provide adequate wound support in elderly, malnourished or debilitated patients or in patients suffering from condition which may delay wound healing, infection, minimal acute inflammatory tissue reaction, localized irritation which skin sutures are left in place for greater than 7 days, suture extrusion and delayed absorption in tissue with poor blood supply, calculi formation in urinary and biliary tracts when prolonged contact with salt solutions such as urine and bile occurs and transitory local infection at the wound site.Infections, erythema, foreign body reactions, transient inflammatory reactions and in rare instances dehiscence are typical or foreseeable risks associated with any suture and hence are also potential complications associated with the barbed suture device.Pdo (polidioxanone) material is essentially absorbed between 182 and 238 days post implantation.However, this specific material has in vivo strength retention for up to six (6) weeks.All sutures are technically ¿foreign substances¿ that the human body has a tendency to reject.Ideally this means the body breaks them down and dissolves them over a period of 3 or 4 months.
 
Event Description
Post-op infection.This case is from health authority.It was reported that the patient was treated with a johnson & johnson surgical knot- free suture for the caesarean section incision suture.About a month plus later, it was found that the wound sewed had been infected.The thread tip of the suture had been extruded out of the abdominal incision, where also had an ulcer.The incision was currently healed after cutting off the extruded thread tip.No additional information could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Type of Device
2V-4 3-0 PDO 24 X 24
Manufacturer (Section D)
SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V.
corredor tijuana rosarito 2000
24702 b, ejido francisco villa
tijuana 22235
MX  22235
Manufacturer (Section G)
SURGICAL SPECIALTIES CORPORATION
corredor tijuana-rosarito 2000 #24702b
ejido francisco villa
tijuana 22235
MX   22235
Manufacturer Contact
luis jauregui giannangelo
247 station drive
suite ne1
westwood, MA 02090
MDR Report Key12558783
MDR Text Key274222327
Report Number3010692967-2021-00041
Device Sequence Number1
Product Code NEW
UDI-Device Identifier20705031207933
UDI-Public(01)20705031207933(20)12(17)250930(10)AAGU334
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/22/2021,09/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberSXPD2B426
Device Lot NumberAAGU334
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/02/2021
Device Age8 MO
Event Location Hospital
Date Report to Manufacturer09/22/2021
Date Manufacturer Received09/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age38 YR
-
-