• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD BHR CURVED CUP INTRODUCER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD BHR CURVED CUP INTRODUCER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 90128257
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that, during a thr surgery, the internal wire tensioner of a bhr curved cup introducer broke off upon impaction of the acetabular cup.No surgical delay was reported to occur due to this incident, and the procedure was concluded with a smith and nephew back up device.No patient harm was reported to occur as a consequence of this malfunction.
 
Additional Manufacturer Narrative
H3, h6: it was reported that, during a thr surgery, the internal wire tensioner of a bhr curved cup introducer (part number: 90128257, lot number: s0801109) broke off upon impaction.No surgical delay was reported, and the procedure was concluded with a smith and nephew back up device.Both pieces remained connected to the impactor and were accounted for.No patient harm was reported.The instrument is for use in treatment.A review of the complaint history for the bhr curved cup introducer was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.Similar complaints have been identified.This will continue to be monitored.As the supplier is no longer in business and there is no traceability within sap addition to the age of the instrument (>10 years old), the manufacturing records for the instrument cannot be reviewed.However, the released instrument involved would have met manufacturing specifications at the time of production.A risk management review was performed.No additional risks were identified as result of the reported event.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.It should be noted that the bhr surgical technique (us) states, "examine instruments for wear or damage before use.While rare, intra-operative instrument breakage can occur.Instruments that have experienced excessive use or force may be susceptible to breakage¿.Without return of the actual devices or further information we cannot further investigate or confirm the reported complaint.If the products or additional information become available in the future, this case will be reopened.Without additional information about this particular case, our investigation remains inconclusive.Based on the results of this investigation, the need for corrective or preventative action is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BHR CURVED CUP INTRODUCER
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12561081
Report Number3005975929-2021-00460
Device Sequence Number11102915
Product Code NXT
UDI-Device Identifier03596010533791
UDI-Public03596010533791
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number90128257
Device Catalogue Number90128257
Device Lot NumberS0801109
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2021
Supplement Date Received by Manufacturer01/28/2022
Initial Report FDA Received Date10/01/2021
Supplement Report FDA Received Date01/28/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/22/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-