Model Number 2217-50-008 |
Device Problems
Device-Device Incompatibility (2919); Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the pinnacle 36 mm liner impactor tips metal collar spins freely and will not tighten to the handle.No further information is available.No surgical delay.
|
|
Manufacturer Narrative
|
Product complaint # = > (b)(4) product lot number: ay1204.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).Investigation summary : no device was received for examination, therefore the reported event could not be confirmed.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|