Model Number SXPD2B426
Health Effect - Clinical Codes Unspecified Infection (1930); Ulcer (2274)
Date of Event 08/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A batch review of the finished good lot and components confirmed there were no quality issues noted throughout the incoming inspection, manufacturing, sterilization, in-process, or final inspection process.To date samples or photos have not been returned for review and/or analysis.Without receiving details of the procedure, placement of the device, culture results and patient health history, a definitive root cause for the reported event of infection cannot be confirmed with certainty adverse effects associated with the use of this device may include, wound dehiscence, failure to provide adequate wound support in closure of the site where expansion, stretching or distension occur, failure to provide adequate wound support in elderly, malnourished or debilitated patients or in patients suffering from condition which may delay wound healing, infection, minimal acute inflammatory tissue reaction, localized irritation which skin sutures are left in place for greater than 7 days, suture extrusion and delayed absorption in tissue with poor blood supply, calculi formation in urinary and biliary tracts when prolonged contact with salt solutions such as urine and bile occurs and transitory local infection at the wound site.Infections, erythema, foreign body reactions, transient inflammatory reactions and in rare instances dehiscence are typical or foreseeable risks associated with any suture and hence are also potential complications associated with the barbed suture device.Pdo (polidioxanone) material is essentially absorbed between 182 and 238 days post implantation.However, this specific material has in vivo strength retention for up to six (6) weeks.All sutures are technically ¿foreign substances¿ that the human body has a tendency to reject.Ideally this means the body breaks them down and dissolves them over a period of 3 or 4 months.
 
Event or Problem Description
Post-op infection.This case is from health authority.It was reported that the patient was treated with a johnson & johnson surgical knot- free suture for the caesarean section incision suture.About four months plus later, it was found that the wound sewed had been infected.The thread tip of the suture had been extruded out of the abdominal incision, where also had an ulcer.The incision was currently healed after cutting off the extruded thread tip.No additional information could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
2V-4 3-0 PDO 24 X 24
Manufacturer (Section D)
SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V.
corredor tijuana rosarito 2000
24702 b, ejido francisco villa
tijuana
MX 
MDR Report Key12561682
Report Number3010692967-2021-00042
Device Sequence Number1479259
Product Code NEW
UDI-Device Identifier20705031207933
UDI-Public(01)20705031207933(20)12(17)250531(10)AAGH543
Combination Product (Y/N)Y
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/01/2021
Report Date (Section F) 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/01/2025
Device Model NumberSXPD2B426
Device Lot NumberAAGH543
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date08/09/2021
Device Age18 MO
Event Location Hospital
Date Report to Manufacturer09/22/2021
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 09/22/2021
Initial Report FDA Received Date10/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age32
-
-