| Model Number |
10430603 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/06/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to siemens that an infant patient had to undergo a second topogram scan (low dose overview scan) due to a system malfunction during the use of the reported somatom definition flash system.Aside from the additional x-ray dose, there was no report of adverse event or injury sustained by the patient due to the necessity of the second scan.This report has been submitted with an abundance of caution.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens completed the technical investigation of the reported event.The root cause of reported issue was attributed to, and was resolved by, replacement of the following parts: cu rot - pn / 10183882.Tx module rotation - pn / 10183874.Rx module rot - pn / 10183880.After replacement of the listed parts, the system was brought back to clinical operation.(b)(4).No general product design issue has been identified.Further investigation within the complaint process is not deemed necessary.
|
| |
|