| Model Number |
9697 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that his blood glucose readings were not being stored in the memory of the contour next meter.During troubleshooting, another testing was performed, which was not stored in the meter's memory.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour next meter for evaluation.No test strips were retuned.The customer service mode was run and no anomalies were found.Three control tests were run using the returned meter with the in-house contour next test strips and in-house contour next control solution to verify meter functionality, which gave a satisfactory performance.Then, the returned meter was tested with the in-house contour next test strips using a blood sample, which gave a satisfactory performance.The blood glucose readings obtained during in-house testing were properly stored in the meter's memory.
|
| |
|