• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US CORAIL2 LAT COXA VARA SIZE 13; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US CORAIL2 LAT COXA VARA SIZE 13; CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 3L93713
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Inadequate Osseointegration (2646)
Date of Event 09/17/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient underwent primary thr on (b)(6) 2018 where a corail pinnacle was utilized.Patient was brought back to theatre on (b)(6) 2020 where it seems there was an attempt to remove the stem for ongoing pain and possible infection.Due to the difficulty of removing the stem, the stem was left in situ and the head exchanged (i do not have any additional details, please check to see whether there was a product complaint submitted at this time) patient has now presented on the gold coast with ongoing pain.A decision was made to open the hip and remove the stem (as it appears on x ray that the stem may be loose).On opening the hip a frozen section was taken and shows the hip to be infected.All implants were removed and a cemented stem loosely cemented in situ and pinnacle liner also loosely cemented in - both to act as a temporary spacer.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that the affected side involved in this event was the right side.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot a worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 LAT COXA VARA SIZE 13
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12562424
Report Number1818910-2021-21701
Device Sequence Number1240669
Product Code LZO
UDI-Device Identifier10603295168904
UDI-Public10603295168904
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K190344
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3L93713
Device Catalogue Number3L93713
Device Lot Number5254195
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2021
Supplement Date Received by Manufacturer10/29/2021
12/08/2021
Initial Report FDA Received Date10/01/2021
Supplement Report FDA Received Date11/03/2021
12/09/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +1.5; CORAIL2 LAT COXA VARA SIZE 13; PINN CAN BONE SCREW 6.5MMX40MM; PINN MAR +4 10D 36IDX58OD; PINN SECTOR W/GRIPTION 58MM; ARTICULEZE M HEAD 36MM +1.5; CORAIL2 LAT COXA VARA SIZE 13; PINN CAN BONE SCREW 6.5MMX40MM; PINN MAR +4 10D 36IDX58OD; PINN SECTOR W/GRIPTION 58MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-