• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMPL TWIST MP-1 3.75 MM 1 1.5 MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMPL TWIST MP-1 3.75 MM 1 1.5 MM; DENTAL IMPLANT Back to Search Results
Model Number 2120
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).
 
Event or Problem Description
It was reported that the implant body cannot be removed from the fixture mount.The implant was removed and another implant was placed.Tooth site # 30.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
One impl twist mp-1 3.75 mm 1 1.5 mm (2120) and mount and packaging were returned for investigation.Visual evaluation of the as returned product identified signs of use.Pre-existing condition as noted on the per was unknown.The implant had been placed on tooth #30 (universal) and was removed the same day.The returned product was functionally tested.The mount was not able to disengage from the implant.X-ray or picture image was not provided.Appropriate documentation was reviewed.Dhr review was completed for the subject lot number 2020080603.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review by lot number (2020080603) was performed for similar event and no other similar complaint identified.Based on the available information, device malfunction has occurred and the reported event was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TWIST MP-1 3.75 MM 1 1.5 MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key12562711
Report Number0002023141-2021-02746
Device Sequence Number13731493
Product Code DZE
UDI-Device Identifier00889024013261
UDI-Public(01)00889024013261(17)250901(10)2020080603
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K013494
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/01/2025
Device Model Number2120
Device Catalogue Number2120
Device Lot Number2020080603
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2021
Supplement Date Received by Manufacturer03/02/2022
Initial Report FDA Received Date10/01/2021
Supplement Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-