|
ETHICON INC. PRLNE BLU 4X30IN 6-0 D/A BV-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
|
Back to Search Results |
|
| Model Number |
M8709 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
09/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts are being made to receive a device.To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent attempts are being made to receive additional information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure what was the tissue condition, i.E., normal or thin, calcified, fragile, diseased? were all four reported sutures (m8709, lot rgbdsc) breaking during initial procedure.When the rubber shod was being applied to each suture? please clarify.When did the bleeding occur? were there any other intra-operative complications which could cause the bleeding? other relevant patient comorbidities/concomitant medications? how was the intra-op bleeding treated? does the patient require of 2 units of blood due to suture breakage occurred when the rubber shod was being applied to the suture? what is the physician¿s opinion as to the etiology of or contributing factors to these events? what is the patient's current status? if applicable, will product be returned? if so, please provide the return date and tracking information." if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate events were submitted via 2210968-2021-09091, 2210968-2021-09093 and 2210968-2021-09094.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a femoral tibial bypass procedure on (b)(6) 2021 and the suture was used to suture vascular tissue.During the procedure, the suture was breaking mid strand while suturing vascular tissue.Some of the sutures broke when the rubber shod was being applied to the suture.Another like device of a different lot was used to complete the procedure.Patient experienced excessive bleeding requiring two units of blood.Any additional consequences are unknown.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Date sent to the fda: 10/21/2021.A manufacturing record evaluation was performed for the finished device batch rgbdsc, m870956 and no non-conformances were identified.Additional h6 component code: g07002 ¿ reported condition not confirmed.Actual sample: an empty opened folder and a suture piece of product code m8709 were returned to ethicon inc for analysis with the packaging opened.The product code is multistrand and each packet contains four strands double armed.In order to evaluate the conditions of the returned sample, the needles were not received for evaluation and damages at the surface suture were observed.In addition, the ends were observed cut.After further evaluation crimping marks were observed on the surface suture possibly from a surgical instrument.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.Functional test was performed, and the tensile strength result were above the minimum requirements.Rep samples: an opened folder with three strands double armed of product code m8709 were returned for analysis with the packaging opened.The product code is multistrand and each packet contains four strands double armed.In order to evaluate the conditions of the returned samples, the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand to detect any issue related to damaged or breakage suture and no defects were observed during evaluation.Functional test was performed, and the tensile strength result were above the minimum requirements.Additional information was requested, and the following was obtained: please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure.78yr old female.No weight or bmi was given.What was the tissue condition, i.E., normal or thin, calcified, fragile, diseased? -tissue was normal.Were all four reported sutures (m8709, lot rgbdsc) breaking during initial procedure when the rubber shod was being applied to each suture? please clarify.When did the bleeding occur? -uncertain if shod was applied to each suture.Were there any other intra-operative complications which could cause the bleeding? - no, other reasons for bleeding.The bleeding occurred when the blood flow was re-opened.Needle hole bleeding.Other relevant patient comorbidities/concomitant medications? none known.How was the intra-op bleeding treated? interrupted sutures and everest does the patient require of 2 units of blood due to suture breakage occurred when the rubber shod was being applied to the suture? - unknown/unsure what is the physician¿s opinion as to the etiology of or contributing factors to these events? - physician had no opinion.What is the patient's current status? - patient went home.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|