| Model Number |
TRX-400 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "ruptured device." the device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: broken shell, red particles on the surface of the shell, underweight.A microscopic analysis was performed which identify sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: - a sharp broken on posterior assessed as unidentified (tear) opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "ruptured device." the device has been explanted.
|
| |
|
Search Alerts/Recalls
|