| Model Number |
TRLP-490 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/20/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: broken observed on posterior side assessed as surgical impact opening (shell thickness was within specification).Additional observations: creases were observed.Wear abrasion observed.Stress marks observed.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported side rupture.The device has been explanted and replaced.
|
| |
|