| Model Number |
PM3210 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the battery longevity estimate was 2 years in dec, 2020, 11.7 months in june, 2021 and less than 3 months to the elective replacement indicator (eri) in sep, 2021.The device had been implanted for 8 years.No anomalies were noted throughout the life of the device but there was a concern of battery overestimation of longevity.The device was to be explanted once eri was reached.The patient was stable before, during and after the event.
|
| |
|
Event or Problem Description
|
|
New information received notes the device was explanted after reaching the elective replacement indicator (eri).
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of incorrect measurement was not confirmed.Device image indicated that autocapture feature was enabled and field data shows device operated in rate responsive mode.When automatic settings are programmed, such as autocapture and rate responsive mode, the device output would adjust itself to accommodate the patient¿s needs for pacing and thus affect the estimated longevity of the device.Analysis performed in nominal settings, including testing at various pulse amplitudes, did not find any anomalies.The device was implanted for 8.54 years which exceeded the device¿s 7.91 year projected longevity and is considered normal battery depletion.
|
| |
|
Event or Problem Description
|
|
New information notes the patient was stable before during and after the explant procedure for the elective replacement indicator.
|
| |
|