| Model Number |
ML02-0115 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Arrhythmia (1721); Sore Throat (2396)
|
| Date of Event |
09/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The disinfectant spill was cleaned up using a medivators disinfectant spill clean-up kit.No additional reports of adverse events or harm were received.The bottle subject of the event was not returned for evaluation.The exact cause of the reported bottle damage could not be determined; however, is likely the result of rough handling during shipping or by user facility personnel.Medivators will continue to monitor for similar events to ensure the product continues to perform as expected.
|
| |
|
Event or Problem Description
|
|
A user facility reported that a bottle of medivators rapicide pa part a had cracked and leaked into the facility's storage cabinet.When attempting to identify the source of the leak, a user inhaled the fumes and reported experiencing throat discomfort and an irregular heartbeat.The user was evaluated on-site at the facility; a chest x-ray was obtained but no additional medical intervention was reported.
|
| |
|
Search Alerts/Recalls
|