| Model Number |
4028210 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation identified steps in the preparation for coating process which can lead to lack of paint adhesion resulting in paint chipping.Additionally, collisions of the light heads and use of unapproved cleaners can further contribute to paint chipping from the light system.The damage to the painted surfaces usually does not develop suddenly but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The device involved in this event passed it's expected lifetime of 10 years.An offer will be made to the customer to replace the damaged lights.The surface coating process has been updated to include additional testing and verification.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
Customer reported that the paint is falling down from the surgical light.The device was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint (b)(4).
|
| |
|
Search Alerts/Recalls
|