H3, h6: the reported device, intended for use in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the exterior of product and no physical damage was observed.Product failed functional testing with footswitch error.Cause of error is a corroded actuator on right pedal which is stuck in the ¿on¿ position.The complaint was confirmed and the root cause has been determined to be corrosion of right pedal actuator.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
|