• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF ANAT BRG RT MD SIZE 3 PMA; OXFORD HXLPE BEARINGS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF ANAT BRG RT MD SIZE 3 PMA; OXFORD HXLPE BEARINGS Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994)
Date of Event 09/27/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Initial report.Report source, foreign - event occurred in (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation due to hospital policy.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient is having a primary total hip replacement and a revision of an oxford unicompartmental knee replacement on the same side (right).The patient had issues of pain and inflammation on the right knee, hence he decided to have it revised at the same time as his hip.
 
Event or Problem Description
It was reported that the patient is having a primary total hip replacement and a revision of an oxford unicompartmental knee replacement on the same side (right).The patient had issues of pain and inflammation on the right knee, hence he decided to have it revised at the same time as his hip.
 
Additional Manufacturer Narrative
(b)(4).This final report is being submitted to relay additional information.Complaint summary: product has not been returned for evaluation.X-rays or medical notes have not been provided.The investigation has been limited to the information provided, a review of the device history records, and a complaint history search.Review of the device history records identified no deviations or anomalies during manufacturing that could be related to the reported event.Review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported item and lot combination.This device is used for treatment.Not enough information has been provided to determine if all implants are compatible.These part and lot combinations are not associated with any recalls at the time the search was conducted.The likely condition of the device when it left zimmer biomet is conforming to specification.The root cause of the reported event is likely to be disease progression, which is not considered device related.No corrective action required at this time as the root cause of the reported event is not device related.If any further information is found which would change or alter any conclusions or information, a supplemental will be submitted accordingly.Zimmer biomet will continue to monitor for trends.H3 other text : hospital policy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF ANAT BRG RT MD SIZE 3 PMA
Common Device Name
OXFORD HXLPE BEARINGS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key12566960
Report Number3002806535-2021-00435
Device Sequence Number16419050
Product Code NRA
UDI-Device Identifier05019279786213
UDI-Public05019279786213
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/24/2019
Device Model NumberN/A
Device Catalogue Number159575
Device Lot Number3427892
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2021
Supplement Date Received by Manufacturer01/19/2022
Initial Report FDA Received Date10/04/2021
Supplement Report FDA Received Date01/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/24/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
Patient SexMale
Patient Weight73 KG
-
-