| Model Number |
393-800 |
| Medical Device Problem Codes |
High Readings (2459); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the complaint the reference electrode membrane lot 2944 was leaking blood in the measuring chamber and the customer reported the na result was too high.The discrepant measurement for na was 156mmol/l.(12:27am (b)(6) 2021) and the comparison measurement for na 135mmol/l (12:26am (b)(6) 2021).Calibrations did not fail until 5:30am (b)(6) 2021 of that morning.Failed parameters currently were ph, na, ca and k.Next calibration was good.The on-board schedule qc was run at 6:08am, qc level 3 and failed ph, na and ca.The qc level 3 was re-run and passed all parameters.The next on-board scheduled qc was run at 8:00am, qc level 2 failed ph, k and na.These parameters were now locked until a pathology staff could go to the instrument to perform maintenance and fix the problem.Customer went to instrument (abl800 flex analyzer) and found blood leaking from the reference electrode measuring chamber.Customer cleaned the blood from the instrument and remembraned the reference electrode at 10:30am and ran qc level 2 and level 3 and passed.
|
| |
|
Additional Manufacturer Narrative
|
|
The radiometer investigation has shown that the product failed due to leakage into measurement chamber around reference membrane.The measurement chamber was cleaned and the reference membrane replaced.The investigation is ongoing.
|
| |
|
Additional Manufacturer Narrative
|
|
Preliminary investigation: the membrane and the reference unit were replaced.The preliminary radiometer investigation of the described event could not confirm the failure and the root cause of the leakage issue is still unknown.Radiometer investigation will continue.
|
| |
|
Additional Manufacturer Narrative
|
|
The radiometer investigation is finalized and the root cause is concluded to be 'supplied materials outside specification.'.
|
| |
|