OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-3
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/08/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to omsc for evaluation.Omsc checked the subject device and found that the reported phenomenon could be duplicated.The exact cause of the reported event could not be conclusively determined at this time.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the gastrointestinal surgery, the subject device was used with foot switch mh-317.The user stepped on the mh-317 to exhaust smoke, but when the intra-abdominal pressure became +5 from the set value, the pinch valve kept opening without stepping on the mh-317.The abdominal pressure dropped.After the trouble occurred, the procedure was continued by manually operating the cock of the trocar without using mh-317, and completed.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was returned to omsc for evaluation.Omsc checked the subject device and found that the reported phenomenon could be duplicated.The manufacturing record was reviewed and found no irregularities.Omsc judged that the pinch valve kept opening and the abdominal pressure dropped because the internal mechanism of the foot switch failed.The caused of foot switch failure could not be identified, but it was presumed that load more than expected was applied on the pedal part and it was damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12567086
MDR Text Key274446532
Report Number8010047-2021-12636
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
PMA/PMN Number
K014166
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received10/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-