• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH DETACHABLE FLEX SHAFT; CHISEL, SURGICAL, MANUAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH DETACHABLE FLEX SHAFT; CHISEL, SURGICAL, MANUAL Back to Search Results
Model Number 2107-2200
Medical Device Problem Codes Fracture (1260); Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was used to drill the screw not in reverse, the spring part was broken apart.No remnant was found in the body on x-ray, but it is unsure.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a trident driver shaft was reported.The event was confirmed via evaluation of the returned device.Method & results: product evaluation and results: visual inspection: visual inspection of the returned device indicated that the outer layer of coils of the flexible shaft have unwound and the coiled wires of the shaft near the tip have fractured.Material analysis: review of the device with a materials analysis engineer indicated that the flexible drill shaft fractured in overload.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there has been 1 other similar event for the lot referenced also relates to crack/fracture.Conclusions: it was reported that the flexible drill shaft fractured during surgery.Visual inspection of the returned device indicated that the outer layer of coils of the flexible shaft have unwound and the coiled wires of the shaft near the tip have fractured.Review of the device with a materials analysis engineer indicated that the flexible drill shaft fractured in overload.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was used to drill the screw not in reverse, the spring part was broken apart.No remnant was found in the body on x-ray, but it is unsure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DETACHABLE FLEX SHAFT
Common Device Name
CHISEL, SURGICAL, MANUAL
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key12567163
Report Number0002249697-2021-01654
Device Sequence Number4053462
Product Code LZO
UDI-Device Identifier07613327212327
UDI-Public07613327212327
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161569
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2107-2200
Device Catalogue Number2107-2200
Device Lot NumberBSKCA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/10/2021
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date10/04/2021
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/04/1999
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age70 YR
Patient SexFemale
-
-