It was reported that during a cryo ablation procedure, a backup console was used and the system notice indicating that the refrigerant delivery path was obstructed, reoccurred.The case was aborted while the patient was under general anesthesia.No patient complications have been reported as a result of this event.
|
|
Product event summary: the data files were returned and analyzed.The data files showed system notice 50012 (the refrigerant delivery path is obstructed) and 50003 (the pressure is low in the system) were both received during the case on multiple occasions.The data files also showed at least nine applications were performed with a non-returned afapro28 balloon catheter with lot 13261 on the date of the event.In conclusion, the system notice of 50012, indicating that refrigerant delivery path is obstructed, and the system notice of 50003, indicating the pressure is low in the system, are not likely to cause or contribute to a death or serious injury; however, the risk of the patient being under general anesthesia without full therapeutic effect is the adverse event being reported.The decision to abort the procedure without use of alternate therapy was based upon the medical judgment of the physician.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|