| Model Number |
IPN914229 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
| |
|
Event or Problem Description
|
|
Shaft of pgac300 broke towards the distal tip of the product when armed and inserting into the abdomen.
|
| |
|
Event or Problem Description
|
|
Shaft of pgac300 broke towards the distal tip of the product when armed and inserting into the abdomen.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The dhr was reviewed.All operations and documentation were completed and acceptable.The device was manufactured on september 16, 2020.Root cause: the cause of the damaged device is undetermined.It is likely that the device was damaged due to excessive force applied while inserting or guiding the needle through the abdomen.An "instruction for use" (ifu) is placed in each min igrip device sales unit.This is a requirement listed on the minigrip device sales unit print, part number tfx-001091 rev 2.Placement of ifu in each sale unit is mandatory per work instruction, wl-cr-9.4 packaging sterilized product.Corrective actions implemented: root cause is unknown , and no corrective actions are necessary at this time.
|
| |
|