Model Number 3341152
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Capsular Contracture (1761); Local Reaction (2035); Paresthesia (4421)
Date of Event 10/14/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old caucasian female patient who underwent a primary breast augmentation procedure with mentor memoryshape breast implant 345cc gel breast prostheses developed baker grade iii capsular contracture in both of her breasts post implantation.Additionally, it was reported that the patient was suffering from bilateral breast pain, numbness, swelling, burning sensation, and the breasts were hot to the touch.Lastly, the patient experienced movement of her breast prostheses while lying down.As a result, the patient underwent bilateral explantation with no replacement breast prostheses on (b)(6) 2021.This medwatch report is for the patient¿s right breast prosthesis.
 
Additional Manufacturer Narrative
On 09-jan-2022, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the patient experienced displacement of the breast implant.In addition, the patient developed capsular contracture and experienced numbness and swelling.The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During the visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Mentor concluded that the capsular contracture in the patient¿s breast was the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsules, and effusion to pathology for examination.Paresthesia is an abnormal sensation such as tingling, tickling, pricking, numbness, or burning of a person's skin with no apparent physical cause.This includes increased or decreased sensitivity or sensation.The manifestation of paresthesia may be transient or chronic.Swelling is a temporary normal post-operative condition.Depending on the presentation, delayed swelling may require additional workup by the surgeon.Swelling is a known complication associated with these devices and is referenced in our current product insert data sheet.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Implant displacement/migration is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 22-dec-2021, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYSHAPE BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
gabriel alfageme
3041 skyway circle north
irving, TX 75038
9497898687
MDR Report Key12569421
Report Number1645337-2021-10989
Device Sequence Number3686811
Product Code FTR
UDI-Device Identifier00081317005193
UDI-Public00081317005193
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P060028
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2021
Device Model Number3341152
Device Catalogue Number3341152
Device Lot Number7345415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/10/2021
Supplement Date Received by Manufacturer12/22/2021
01/09/2022
Initial Report FDA Received Date10/04/2021
Supplement Report FDA Received Date01/06/2022
01/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-