BOSTON SCIENTIFIC CORPORATION 2CM PERIPHERAL CUTTING BALLOON; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA Back to Search Results
Model Number 24630
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that balloon detachment occurred and was removed through a snare.The 70-90% stenosed target lesion was located in the mild to moderately tortuous and a very mild, intimal fibrosis calcified fistula.A 8.00mm/2.0cm/90cm peripheral cutting balloon was selected for use.During withdrawal, resistance was felt.When the device reached the end of the sheath, it was noted that the balloon was detached from the catheter.The catheter came out through the sheath over the wire and was removed altogether leaving the balloon within the fistula.The physician was able to access a new location site and removed the balloon through snaring.No further complications were reported and patient was well post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2CM PERIPHERAL CUTTING BALLOON
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINA
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12569573
Report Number2134265-2021-12250
Device Sequence Number872129
Product Code PNO
UDI-Device Identifier08714729813804
UDI-Public08714729813804
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K151253
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2022
Device Model Number24630
Device Catalogue Number24630
Device Lot Number0026088693
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/15/2021
Initial Report FDA Received Date10/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-