• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE SECTOR II CUP 62MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE SECTOR II CUP 62MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-22-062
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Naturally Worn (2988); Noise, Audible (3273)
Health Effect - Clinical Codes Fall (1848); Foreign Body Reaction (1868); Joint Laxity (4526)
Date of Event 09/21/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient, with spinal fusion, had bilateral thas and a few months after surgery the left one felt as though it was subluxing.9 months post-op he had a fall and the liner was found to be dissociated from the cup on x-ray.He now felt grinding of the femoral head on the cup.The hip was revised, metallosis was observed as the ceramic head was wearing the shell.The liner, head and shell were removed and new revision components were implanted.Doi: 2021, dor: (b)(6) 2021, unknown side.
 
Event or Problem Description
Additional information received indicated that there was no surgical delay.Affected side was the left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary examination of the returned device and provided x-ray images found evidence to support disassociation of the liner from the cup while implanted.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot a worldwide complaint database search found no additional related reports against the provided product code/lot number combination.Based on the inability to find any additional related reports against the provided product code/lot number combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE SECTOR II CUP 62MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key12570570
Report Number1818910-2021-21856
Device Sequence Number17016168
Product Code LPH
UDI-Device Identifier10603295009870
UDI-Public10603295009870
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-22-062
Device Catalogue Number121722062
Device Lot NumberC83403
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2021
Supplement Date Received by Manufacturer09/29/2021
10/22/2021
12/08/2021
Initial Report FDA Received Date10/04/2021
Supplement Report FDA Received Date10/13/2021
11/11/2021
12/09/2021
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX NEUT 36IDX62OD; DELTA CER HEAD 12/14 36MM +1.5; PINNACLE SECTOR II CUP 62MM; ALTRX NEUT 36IDX62OD; DELTA CER HEAD 12/14 36MM +1.5; PINNACLE SECTOR II CUP 62MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
Patient Weight94 KG
-
-