| Model Number |
G-SD30-1550 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(b)(6) field service technician submitted complaint alleging that the cotter pins that holds the wheel assembly together on the integrity 270 sterilizer carriage rack were coming loose.If a wheel comes off the rack there is a risk of the rack tipping and contents falling.
|
| |
|
Additional Manufacturer Narrative
|
|
There is no new action implemented for this event.Evaluation concluded: 1) the instrument loading rack in/out test (repeated test), equivalent to seven (7) years of sterlizer operation was conducted and it was concluded that the wheels do not fall off under normal operation.As the facility reported the event, the customer uses multiple transfer carts and loading racks and it is likely the transfer cart/loading rack came into contact with one anotehr while moving where external force may have contributed to snap the pin, where the wheel came off.2) there is no risk per the risk assessment/evaluation.3) there were no other complaints/reports relating to this event.
|
| |
|
Event or Problem Description
|
|
Skytron field service technician submitted complaint alleging that the cotter pins that holds the wheel assembly together on the integrity 270 sterilizer carriage rack were coming loose.If a wheel comes off the rack there is a risk of the rack tipping and contents falling.
|
| |
|