Model Number URF-P6
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2021
Event Type  malfunction  
Event Description
An olympus employee observed a deviation in cleaning, handling and incorrect reprocessing of uretero-reno fiberscopes by the tiger team.The scope was left on the case for several hours prior to transport to sterile processing department (spd) for manual cleaning.There was no harm, infection or user injury reported due to the event.This event is related to patient identifiers: (b)(6).
 
Manufacturer Narrative
The olympus employee observed dirty scopes were left in the case tray waiting to be transported to sterile processing department (spd) for cleaning for over an hour.The scopes used in the morning were not transported to spd for manual cleaning until the end of the day.The olympus employee documented the cleaning, disinfection, and sterilization process for the devices.The customer was instructed to follow all olympus reprocessing procedures.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Manufacturer Narrative
The supplemental report is submitted to provide the result of the legal manufacturer¿s investigation.Updates to section.The device history record for the subject device was reviewed and it was verified that the device was manufactured in accordance with documented specifications.The legal manufacturer performed an investigation.A conclusive root cause was not identified.The investigation determined that the user facility did not train reprocessing staff sufficiently on handling and reprocessing in accordance with information for use.(ifu).Additionally, it was confirmed that there was cleaning delay, wrong handling and transportation.The scope was on the table with other instruments stacked on it post procedure scope was left on the case cart for several hours prior to being transported to spd for manual cleaning.Ess recommended the user to follow all the olympus reprocessing process, described in on track form, and reprocessing ifu.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: precautions: an insufficiently cleaned, disinfected or sterilized endoscope and/or accessories may pose an infection control risk to the patients and/or operators who contact them.If the endoscope and accessories used in the patient procedure are not immediately cleaned after each patient procedure, residual organic debris will begin to dry and solidify, hindering effective removal and reprocessing efficacy.Preclean the endoscope and the accessories at the bedside immediately after each patient procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key12571634
MDR Text Key280938498
Report Number8010047-2021-12664
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-