MAUDE Adverse Event Report: JJGC S.A. SELF-DRILLING ANCHORAGE IMPL MB TI 1.3X7; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JJGC S.A. SELF-DRILLING ANCHORAGE IMPL MB TI 1.3X7; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Model Number 109.487
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.  the fracture of a dental implant is a known long-term complication of endosseous dental implants.The manufacturers trend analysis confirms that the reported implant fracture rate associated with its dental implants is low and remains consistent with the experience as documented in the literature.There are different reasons why implants break and this problem has been illuminated by various authors (e.G.Maeglin 1988, beumer 1989, tetsch 1991, worthington 1992).Beumer and lewis 1989 report the following causes: production / design flaws, inadequate fitting of the suprastructure, occlusal overload, bruxism, bone resorption around the implants, insufficient axle alignment of the implants, trauma, biomechanical causes, and design of the suprastructure.
 
Event or Problem Description
The clinician reports the implant was inserted in the patient's mouth.On (b)(6) 2021, fracture of the implant was verified.The device was forwarded to the manufacturer.There were no patient operative or post-operative complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF-DRILLING ANCHORAGE IMPL MB TI 1.3X7
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key12571728
Report Number3008261720-2021-07012
Device Sequence Number3419949
Product Code OAT
UDI-Device Identifier07898237567532
UDI-Public07898237567532
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K102769
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 10/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/05/2026
Device Model Number109.487
Device Catalogue Number109.487
Device Lot NumberFHL32
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/04/2021
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-