| Model Number |
7133 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
09/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a shaft leak and clotting occurred.The target lesion was located in the mid left anterior descending artery.A 3.00mm x 15mm emerge balloon catheter was advanced for dilatation.The balloon was used for multiple inflations; however, upon another inflation, a leak was noticed in the mid shaft.The patient became symptomatic and a clot was noticed.Manual thrombectomy was performed and the procedure was completed with another of same device.No further patient complications were reported and the patient recovered.
|
| |
|
Event or Problem Description
|
|
It was reported that thrombosis occurred.The target lesion was located in the mid left anterior descending artery.A 3.00mm x 15mm emerge balloon catheter was advanced for dilatation.The balloon was used for multiple inflations; however, upon another inflation, a leak was noticed in the mid shaft.The patient became symptomatic and a clot was noticed.Manual thrombectomy was performed and the procedure was completed with another of same device.No further patient complications were reported and the patient recovered.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: returned product consisted of an emerge mr balloon catheter.The device was visually and microscopically examined.There were numerous kinks to the hypotube of the device.There was contrast and blood in the inflation lumen and the balloon.There was blood in the guidewire lumen.The balloon was loosely folded.The device was prepped with an inflation device filled with water and connected to the inflation port to inflate the device.The device failed to inflate to the rated burst pressure.At 30.1cm from the tip of the device for a length of 2mm there was a tear in the inflation lumen causing a leak in the device.Product analysis confirmed reported shaft leak as the inflation lumen was torn.
|
| |
|
Search Alerts/Recalls
|