Device expiration date: unknown.Device manufacture date: unknown.Investigation summary: our quality engineer inspected the sample and photographs submitted for evaluation.Bd received one used unit and two photos.During the visual examination, it was observed that the unit was missing the clamp the reported issue was confirmed.No damage was present on the extension tubing indicating the defect most likely occurred during the manufacturing process.There is an automated vision system in place that inspects the products during manufacturing and preventative maintenance is regularly performed to mitigate the risk from this type of defect.A device history record review could not be performed as the lot number is unknown.The appropriate personnel have been notified and we appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.Investigation conclusion: the reported defect has been confirmed based on photo and physical evidence.The root cause has been linked to the manufacturing.
|