• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/10/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Testing of actual/suspected device : a getinge field service engineer (fse) was dispatched to evaluate the iabp and was unable to reproduce the reported issue.The fse ran device on battery power during checks, looked for any physical damage to wires and cables, but could not find any damage on internal components.Unit ran all weekend without any issues.The fse completed a pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) shut down.No patient harm, serious injury or adverse event was reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, h2, h4, h6 (investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1 (contact person).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12572887
MDR Text Key274758744
Report Number2249723-2021-02249
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received06/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2009
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-