It was reported that, after a primary tkr surgery, had been performed on 2016, one lgn ps high flex xlpe sz 7-8 11mm and one 35mm resurfacing patella appear to trigger pain and avascular necrosis (avn).The problem was treated with a revision surgery on (b)(6) 2021, where an 7/8 12mm high flex was implanted.Current health status of patient is unknown.
|
|
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, smith and nephew has not received the explanted device and/or adequate materials to fully evaluate the complaint.Without the requested information, the clinical root cause of the reported event could not be fully assessed or concluded.The patient impact beyond the reported symptoms and revision could not be determined, as the current patient status is unknown.If additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, joint tightness, material in use, patient reaction, loss of sterility and/or patient condition.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|